Customization: | Available |
---|---|
Type: | Insulated |
Conductor Type: | Stranded |
Still deciding? Get samples of US$ 20/Meter
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
The 24 Wire MRI Cable With 12 Coax 32AWG And 12x 26AWG Signal Wire is widely used in MRI Machine repair.We also provide other spec MRI cable with different coax numbers and signal wire numbers,such as 12 coax +20 signal wire,16 coax+16 DC wires…..It maybe used in 24-channel head coil for MR imaging
And we are experienced in custom MRI cable for MRI rf coils
Regular MRI compatible Configuration List:
Item | Unit | Specified Value | ||||||||||||||||
32AWG Coaxial (A) | 26AWG Wire(B) | |||||||||||||||||
Number of cores | - | 12 | 12 | |||||||||||||||
Inner Conductor | Material | - | Silver plated Copper wire | |||||||||||||||
Stranding | No./mm | 1/0.20 | 19/0.10 | |||||||||||||||
Diameter(Nom.) | mm | 0.20 | 0.50 | |||||||||||||||
Insulation | Material | - | FEP | FEP | ||||||||||||||
Thick.(Nom.) | mm | 0.20 | 0.15 | |||||||||||||||
Diameter | mm | 0.60±0.02 | 0.80±0.02 | |||||||||||||||
Color | - | Nature | See Fig.1 | |||||||||||||||
Braid Shielding | Material | - | Silver plated Copper wire | - | ||||||||||||||
Strand Dia. (approx.) | mm | 0.08 | ||||||||||||||||
Dia.(approx.) | mm | 0.92 | ||||||||||||||||
Coverage | % | 90% | ||||||||||||||||
Sheath | Material | - | FEP | |||||||||||||||
Thick.(Nom.) | mm | 0.05 | ||||||||||||||||
Diameter | mm | 1.18±0.03 | ||||||||||||||||
Color | - | See Fig.1 | ||||||||||||||||
Assembly Diameter(Nom.)1 | mm | 3.3(See Fig.1) | ||||||||||||||||
Assembly Diameter(Nom.)2 | 5.88 | |||||||||||||||||
Tape1 | Material | - | PTFE | |||||||||||||||
Thick.(Nom.) | mm | 0.05 | ||||||||||||||||
Diameter | mm | 6.08 | ||||||||||||||||
Braid Shielding | Material | - | Tinned copper wire | |||||||||||||||
Dia.(approx.) | mm | 0.12 | ||||||||||||||||
Coverage | % | 85% | ||||||||||||||||
Tape2 | Material | - | PTFE | |||||||||||||||
Thick.(Nom.) | mm | 0.05 | ||||||||||||||||
Diameter | mm | 6.76 | ||||||||||||||||
Jacket | Material | - | TPU | |||||||||||||||
Thickness (Nom.) | mm | 0.85 | ||||||||||||||||
Diameter | mm | 8.50±0.20 | ||||||||||||||||
Color | - | Gray/Custom |
• Operating voltage:30V
• Test voltage:AC 1.0kV/1min
• Temperature range:-40°C ~+ 80°C
• Insulation resistance B(20°C):≥ 100MOhm x km
• Characteristic impedance:50±3Ω
• Delay:4.9ns/m
• VSWR:≤1.3(12~300MHz)
• Attenuation typ.(dB/100m):12MHz≤21.7,63MHz≤49.5,
123MHz≤69.1,135MHz≤72.5,300MHz≤108.8 ,350MHz≤155.2,
400MHz≤175.1,450MHz≤190.2,500MHz≤210.1